KM9106 Operators Manual
46
To access timed purge enter the following :-
Enter the code 0000 at the following display :-
ENTER SERVICE CODE
-- -- -- -- > 0 0 0 0 < -- -- ---
The following screens will be displayed, enter NO to both :-
CALIBRATE SENSOR
NO
PRINT MEMORY ?
NO
The purge sequence now begins, by using the arrow keys select YES to
‘SET PURGE ?’ and press .
SET PURGE ?
YES
Enter the purge duration and similarly the purge interval by changing each character.
Note the line below the zero indicates the number to be changed. Enter when correct.
If a time of less than the minimum or greater than the maximum then these will set as
the default values.
PURGE DURATION 05
<05 . . . . . . . . . . . . 09>
PURGE INTERVAL 30
<09 . . . . . . . . . . . . 99>
SET/CAL
ENTER
SET/CAL
ENTER
ENTER